Adjustments
I/Q Impairment Calibration
4-42
I/Q Impairment Calibration
Description
This calibration sets internal calibration array values to minimize errors in the I/Q adjustments (I/Q gain, I
offset, Q offset, and quadrature skew) in the I/Q menu.
NOTE This adjustment assumes the I/Q Gain/Offset/Quadrature calibration has already been
performed.
Required Test Equipment
•None
Procedure
1. Preset the signal generator.
2. Follow the instructions as they appear on the controllers display.