112 Chapter 2
Performance Verification Tests
14. Reference Level Accuracy: Agilent E4401B and E4411B
a. Press Single on the analyzer.
b. Press
Peak Search (or Search).
c. Record the Marker Delta Amplitude reading in Column 5 of
Table 2-21.
d. Add the Actual Attenuator Step to the Analyzer Marker Delta
Amplitude and enter the result as the Test Record entry in the
performance test record.