General Power Meter Functions 2
N1913/1914A EPM Series Power Meters User’s Guide 91
Memory Erase/Secure Erase
The memory erase and secure erase features will erase the battery backed
SRAM, and flash file system. The flash file system includes the power
meter states, cal factor tables, frequency dependant offset tables, and the
secure blank password stored in the EEPROM. Upon completion, the meter
‘s contents will be initialized to the default settings. These features can
only be activated via the front panel.
The memory erase feature can be invoked when you forget the 6- digit
password set during secure blank.
To perform memory erase/secure erase, do one of the followings:
Memory Erase
1 Simultaneously press the fourth softkey from the top and the
key immediately upon power- on (see Figure 2- 55).
2 During memory erase, the splash screen will display “Clearing Memory...”.
When memory erase is complete, “Clearing Memory...done” will be
displayed.
Figure 2-55 Activate memory erase using combo keys
Press simultaneously