1053210A 53220A 53230AMath operationsSmoothing (averaging) 1Selectable 10 (slow), 100 (medium), 1,000 (fast) reading moving average Selectable lter r
1153210A 53220A 53230AMeasurement/IO timeout (nom) no timeout or 10 ms to 2000 s, in 1 ms stepsAuto-level speed Slow mode (50 Hz): 350 ms (typ) Fast
1253210A 53220A 53230AMaximum measurement speed to internal non-volatile memory6: (readings/s)Timestamp N/A N/A 1,000,000Frequency, period, totalize
1353210A 53220A 53230AWarm-up time 45-minutesDisplay 4.3" Color TFT WQVGA (480 x 272), LED backlightUser interface and help languagesEnglish, Ge
1453210A 53220A 53230ABattery (Option 300)Technology Internal lithium ion battery with integrated smart battery monitor & chargerUse for maintaini
15TimebaseStandard TCXOOption 010 Ultra-High Stability OCXOAging 1 (spec) 24-hour, TCAL ±1 °C ± 0.3 ppb (typ) 30-day, TCAL ±5 °C ± 0.2 ppm (typ)
16Basic accuracy 1 = ± [(k * Random Uncertainty) + Systematic Uncertainty + Timebase Uncertainty] Measurement Function1-σ Random UncertaintySystematic
17Accuracy Specications (continued)Measurement Function1-σ Random UncertaintySystematic UncertaintyTimebase Uncertainty26 GHz (Option 106): Optional
181. Apply the appropriate errors detailed for each measuring function. 2. Use Timebase Uncertainty in Basic Accuracy calculations only for Meas
19Resolution enhancement factor (RE) The resolution enhancement (RE) calculates the added frequency resolution beyond the basic reciprocal measurement
2IntroductionFrequency counters are depended on in R&D and in manufacturing for the fastest, most accurate frequency and time interval measuremen
20±(0.2%-of setting + 0.1%-of range) V/s (at threshold point) Denition of Measurement Error Sources and Terms used in Calculations (continued)Phase N
21Ordering InformationModel numbers53210A 350 MHz, 10-digits/s RF Frequency Counter 53220A 350 MHz, 12 digits/s, 100 ps Universal Frequency Counter
22Parameter assumptions: – 53220A – 95% condence – 100 MHz signal, 1 sec gate – AUTO frequency mode – Level: 5 V input signal amplitude – TCXO standa
23Specication (spec)The warranted performance of a calibrated instrument that has been stored for a minimum of 2½ hours within the operating temperat
myKeysightwww.keysight.com/find/mykeysightA personalized view into the information most relevant to you.www.axiestandard.orgAdvancedTCA® Extensions fo
3Input characteristics (nom) 53210A 53220A 53230AChannelsStandard (DC - 350 MHz) Ch 1 Ch 1 & Ch 2Optional (6 or 15 GHz)Ch 2 Ch 3Standard inputs
453210A 53220A 53230AOptional microwave inputs (nom)Frequency rangeOption 106100 MHz - 6 GHzOption 115300 MHz - 15 GHzInputConnectorFront panel precis
553210A 53220A 53230AMeasurement range (nom)Frequency, period (average) measurementsCommonChannels Ch 1 or optional Ch 2 Ch 1, Ch 2 or optional Ch 3Di
653210A 53220A 53230ATime interval A to B, B to ARange9N/A -1 ns to 100,000 s (nom) -0.5 ns to 100,000 s (min)Time interval A or BRange N/A 2 ns to 10
753210A 53220A 53230A6 GHz (Option 106) 15 GHz (Option 115)Pulse/burst frequency and pulse envelope detector (Option 150)12Pulse/burst measurements N/
8Gate, Trigger and Timebase Characteristics53210A 53220A 53230AGate characteristics (nom)GateSource Time, external Time, external or advancedGate tim