Characteristics and Specifications
Characteristics and Specifications
Characteristics and specifications for the N2795A/N2796A Single-ended Active Probes
are shown below. The probe / oscilloscope should be warmed up for at least 20 minutes
before any testing and the environmental conditions should not exceed the probe’s
specified limits.
Electrical Characteristics
Attenuation ratio (@ DC) N2795A: 10:1 +/- 0.5%
N2796A: 10:1 +/- 0.5%
Bandwidth (-3 dB)* N2795A: 1 GHz
N2796A: 2 GHz
Probe Risetime (calculated,
10%-90%)
N2795A: 350 ps
N2796A: 175 ps
Input Dynamic Range N2795A: -8V to +8V (DC or Peak AC)
N2796A: -8V to +8V (DC or Peak AC)
Input Resistance* N2795A: 1 MW +0%, -2.5%
N2796A: 1 MW +0%, -2.5%
Non-destructive Max Input
Voltage
N2795A: -20V to +20V
N2796A: -20V to +20V
Offset Range N2795A: +/- 8V
N2796A: +/- 12 V
DC Offset Error (Output Zero) N2795A: < +/- 1 mV
N2796A: < +/- 1 mV
Flatness (both models) Typical 0.4 dB (100 kHz - 100 MHz)
Typical 0.6 dB (100 MHz - 500 MHz)
Typical 0.8 dB (500 MHz - 1 GHz)
Typical 1.3 dB (1 GHz - 2 GHz)**
Input Capacitance N2795A: 1 pF
N2796A: 1 pF
Probe Noise (referred to input) N2795A: < 1 mVrms
N2796A: < 1 mVrms
Output Impedance N2795A: 50 W
N2796A: 50 W
Internal Power Agilent AutoProbe Interface from Scope
(InfiniiVision and Infiniium models)
*Denotes warranted specification after 20 minute warm-up, all others typical
**N2796A only