Agilent Technologies 34980A User's Guide Page 1

Browse online or download User's Guide for Network switches Agilent Technologies 34980A. Agilent Technologies 34980A User`s guide User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - Agilent Modular Products

Agilent 34980A Multifunction Switch/Measure Unit Data SheetChallenge the Boundaries of TestAgilent Modular Products

Page 2 - Optimized for test systems

10Table 3. Multiplexer selection table—specifications and characteristics 34921A 34922A 34923A 34924A 34925AChannels/configurations 402-wire 702

Page 3 - Measurements you can trust

11Table 3. Multiplexer selection table—specifications and characteristics—continued 34921A 34922A 34923A 34924A 34925AGeneral characteristicsRelayli

Page 4 - Remote access and control

1234980A matrix switch modulesThe 34980A matrix modules are full cross-point matrices that allow you to connect any row to any column. This is a conve

Page 5 - Also Available

13Table 4. Matrix selection table—specifications and characteristics 34931A 34932A 34933A 34934AChannels/configurations dual 4x8 dual 4x16 dual

Page 6

1434980A general-purpose switch modules The 34980A general-purpose switches can be used to route signals or to control other system devices. These sw

Page 7

1534980A RF and microwave switch modulesThe34980AoffersavarietyofRFandmicrowaveswitchmodules—RFmultiplex-ers, SPDT switching from DC to 26.

Page 8 - 34980A multiplexer

1634946A/47A—from DC to 26.5 GHz For applications where you need only a few high-frequency switches, the 34946A and34947Aoffersingle-pole,double-t

Page 9

17Table 6. RF and microwave selection table—specifications and characteristics DC to 3 GHz DC to 26.5 GHz [3] 34941A 34942A 34946A 34947AChannels

Page 10

1834945A/34945EXT microwave switch/attenuator driverThis module allows you to control switches attenuators, and other devices external to the 34980A.

Page 11

1934950A 64-bit digital I/O with memory and counter This module can be used to simulate or detect digital patterns. It has eight 8-bit digital I/O c

Page 12 - . .

If you use automated test equipment for design validation or manufacturing, you now have a cost-effective solution for many test system applications.

Page 13

2034951A 4-channel isolated D/A converter with waveform memory This module has four independent, isolated channels that output DC voltage up to ± 16

Page 14 - 34980A general-purpose

2134952A multifunction module with 32-bit DIO, 2-channel D/A and totalizer The multifunction module offers the flexibil-ity you need for system contro

Page 15 - 34980A RF and microwave

2234959A breadboard module Use this module to create your own custom designs inside the 34980A mainframe. You can control your custom circuits with ac

Page 16

2334980A system specifications and characteristics DMM accuracy ± (% of reading + % of range) Includes measurement error, switching error, and transd

Page 17

24Additional Low Frequency Error for ACV, ACI (% of reading)Frequency AC Filter Slow AC Filter Medium AC Filter Fast10Hz-20Hz 0 0.74 – 20Hz

Page 18 - 34945A/34945EXT microwave

25Scanning measurement rates to bus or memory Direct measurements – direct to I/O Measurement (includes switch, measure time and I/O time) into

Page 19 - 34980A system control modules

26Measurement characteristics with optional internal DMMDC voltageMeasurement method Continuously integrating multi-slope III A-D converterA-Dl

Page 20 - D/A converter with

27Measurement characteristics with optional internal DMM continuedDC operating characteristics [4]Function Digits[5] Readings/s Additional RMS noise

Page 21 - 34952A multifunction module

28General specificationsPowersupply Universal100Vto240V±10%Powerlinefrequency 50–60Hz±10%automaticallysensedPower consumption 1

Page 22 - 34959A breadboard module

29Ordering instructions Mainframe – holds up to 8 plug-in modules 34980A Multifunctionswitch/measuremainframe Comesstandardwith“DMM”option,B

Page 23

Easier signal routing with four 2-wire internal analog buses. You can route your measurements directly to the inter-nal DMM, or you can connect to ex

Page 24

30Screw terminal blockOrdering instructions continued RF and microwave modules 34941A Quad1x450-ohm3-GHzRFmultiplexer 20–SMA Requiresstanda

Page 25 - mode mode mode

31Cables [1] used for direct cable connection to module. some modules require 2 cablesY1135A 1.5m50-pinDsub,M/Ftwistedpairwithoutershieldc

Page 26

For more information on Agilent Technologies’ products, applications or services, please contact your local Agilent office. The complete list is ava

Page 27 - System specifications

4Standard interfaces take the has-sle out of connecting to your PC Standard Ethernet, USB and GPIB inter-faces are included in every mainframe. Use on

Page 28 - Software

5Free BenchLink Data Logger Software to Simplify data loggingTheBenchLinkDataLoggersoftwareforthe34980A provides a convenient way to collect an

Page 29 - Ordering instructions

6Intuitive front panel with self-guiding menusSet up scan listsSee results on bright, multiline displayScan multiple channels, close specified channel

Page 30

7Mix and match 34980A modules to create your own custom configurationThe 34980A mainframe holds up to eight plug-in modules. Mix and match them to cr

Page 31

834980A multiplexer switch modulesThe 34980A multiplexer modules can be used to connect one of many different points to a single point. You can conne

Page 32 - Agilent Solutions Partners

9Multiple multiplexers can connect to the built-in analog buses, allowing you to scan up to 560 2-wire channels or 640 1-wire channels in a single ma

Comments to this Manuals

No comments