Agilent Technologies 34980A User's Guide Page 12

  • Download
  • Add to my manuals
  • Print
  • Page
    / 32
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 11
12
34980A matrix switch modules
The 34980A matrix modules are full cross-
point matrices that allow you to connect
any row to any column. This is a convenient
way to connect multiple test instruments to
multiple points on a device under test.
Choose from the following features:
• Latchingarmaturerelays—300V,1A
• High-speedreedrelays—150V,0.5A
• Configurabledual4x8,dual4x16orquad
4x32 modules
• Single-wireconfiguration
(34933A or 34934A)
• Highdensitymatrixwithautomaticsurge
protection and row disconnect for flex-
ible measurements (34934A)
• Analogbusexpandablerowsto
create larger matrices
(34931A, 32A, 33A)
• Connectionsviastandard50or78-pin
Dsub cables or detachable
terminal block
Each cross-point in the matrix switch has
two wires—a high and a low for the mea-
surement. Or, if you prefer, the 34933A and
34934A can be configured as a single-wire
matrix, increasing the number of channels.
The 34933A also has in-rush resistors on
each column for added protection.
The 34934A also has in-rush protection
resistors, but also has an automatic bypass
switch for flexibility in making low-level
measurements.Rowdisconnectswitches
also reduce the capacitance loading when
combining modules to create larger matri-
cies.
Multiple matrix modules can be combined
through the analog bus or the row expan-
sion kit (34934A only) to create a larger
matrix. The matrix can then be connected to
the internal DMM for easy measurements.
Combine your matrix with a multiplexer
switch to achieve the desired switching
topology and get a lower-cost solution with
better specifications. All the matrix switches
include a relay counter to help predict
when relays need to be replaced. Use
the sequencing feature to easily change
between different cross-point setups.
L
H
L
H
L
H
L
H
921
L
H
922
L
H
923
L
H
924
L
H
LHLHLHLHLH
ABus1 DMM
(MEAS)
ABus2 DMM
(SENS)
ABus3
ABus4
Figure 7. 34932A dual 4x16 armature matrix
Note: The 34933A and 34934A have 100
ohm input protection resistors to limit cur-
rent and protect the reed relays.
Figure 8. 34934A quad 4x32 matrix (1 of 4 matricies shown)
. . .
R 2
R 4
R 3
R 1
C 1 C 32