88 Chapter 3
Programming Command Reference
TESTs Subsystem for MANUAL TEST
BS Release
TESTs:MANual:MEASure:BSRelease
This command releases the mobile from the Test Set. There is no query form of
this command.
NOTE If Option 002 and Option 040 are installed and the mode is set to
ASYNC or GPRS, this command is invalid.
DC Power Voltage
TESTs:MANual:MEASure:DCPower:VOLT <real> [V]
TESTs:MANual:MEASure:DCPower:VOLT?
This command specifies a voltage value of DC Power if the DC Power supply
mode in the Stand-by screen is set to ON. The allowable range is from 3.0 to 11.0
(V) in 0.1 steps. At preset, this is set to 3.0 (V).
GPRS Test End
TESTs:MANual:MEASure:END
This command terminates measurements in the GPRS mode and returns to the
Stand-by screen. There is no query form of this command.
NOTE If the mode is set to SYNC on the Test Set with Option 040, this
command is invalid.
Midamble
TESTs:MANual:MEASure:MIDamble
NONE|TSC0|TSC1|TSC2|TSC3|TSC4|TSC5|TSC6|TSC7
TESTs:MANual:MEASure:MIDamble?
This command specifies a value of Training Sequence Code (TSC), TSC0 to
TSC7, or NONE for the amplitude trigger. At preset, this is set to TSC5.
NOTE For Test Sets without Option 002, or if the mode is set to SYNC,
this command is invalid.
For Test Sets with Option 040, if the mode is set to GPRS, this
command is invalid.
MS Call
TESTs:MANual:MEASure:MSCall
This command executes a call from the mobile. There is no query form of this
command.