Agilent Technologies E6392B User's Guide Page 42

  • Download
  • Add to my manuals
  • Print
  • Page
    / 206
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 41
Chapter 2 41
Using Functions
Testing a Mobile with MANUAL TEST (Asynchronous mode)
2. Using Functions
If you want to supply dc power from the Test Set to the mobile, set the DC Power
field to On and enter your desired voltage value. To do this, use the Universal DC
Power Adapter and an appropriate dc power cable to connect between the AUX
port of the Test Set and the mobile.
Step 7. Press Return, then CONFIGURATION, then Test Setup to obtain the Test Setup:
Test Condition screen shown below.
Step 8. Specify the highlighted fields of the test parameters as shown in the left side of the
above screen.
Step 9. Enter the appropriate limits in the cells of the test item table above.
Step 10. Repeat Step 8 to Step 9 to set the Test Condition screen by replacing DCS1800
with GSM900 in Radio Standard.
Step 11. After you have set all parameters as needed, press the Return, Return again,
MANUAL TEST, then Mode Sync/Async
softkeys to obtain the MANUAL TEST:
Stand-by screen with the asynchronous mode.
Step 12. Set the mobile phone to transmit the bursted RF power.
Step 8 Step 9