98 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz
5 Preventive Maintenance
Instrument Installation
To install or reinstall an instrument in a system, use one of the following procedures.
Required tools
#2 Phillips screwdriver
#2 POZIDRIV screwdriver
system installation guide
Standard rack instrument
NOTE
Detailed information related to rack mounting can be obtained in the
Agilent N8200A Series Synthetic Instrument Modules, Rack Configuration Guide,
N8200-90003.
To install an instrument
Step Notes
1 Slide the instrument gently into the rack.
2 Insert the screws in the rack ears.
Most cables are fairly easy to remove and
reconnect and have no special
considerations (besides making sure you
put the cables back in the right place).
Semirigid cables require more care,
especially when reconnecting them. Make
sure all semirigid cables, on the front and
back of an instrument are fully
disconnected before removing the unit.
3 To reconnect the semirigid cables, carefully
align them before you insert the male
connector.
Do not insert the male pin in at an angle or
you will damage the female connector. RF
connector center pins are very delicate,
and if damaged must be replaced. System
performance may be greatly impaired if
there is a bad RF connector.
4 Turn on system power and restart the system
computer if necessary.