Chapter 3: Testing Performance
To verify threshold accuracy
53
“Oscilloscope Threshold Voltage Setting” in Table 1 on page 55.
Close the Digital Setup dialog.
l Set the horizontal time scale.
Select the horizontal scale setting (near the bottom of the display).
In the Enter Scale dialog, enter either 200 or 500 and then select p.
For MSO/DSO8064A models, enter 500 and then select p.
For MSO/DSO8104A models, enter 200 and then select p.
Select horizontal
scale setting