Chapter 5: Troubleshooting
To install the fan safety shield
73
Default Setup
A Default Setup is provided to ensure that the oscilloscope is in a known state. The
default setup prevents previous setups from interfering with the next test. It also
simplifies the oscilloscope setup procedure. Use the default setup when a procedure
requires it.
Press the Default Setup key to set the oscilloscope to the default state.
To install the fan safety shield
1 Disconnect the oscilloscope power cord and remove the cover.
See Chapter 6, “Replacing Assemblies,” if you need to remove and replace assemblies.
2
Clip the fan safety shield over the outside of the oscilloscope chassis next to the fans.
See See Figure 5-1.
Figure 5-1
Installing the Fan Safety Shield
Page view 72