Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
22
6 Switch on the channel and check the high level at the following settings
(Program the Low Level at least 500 mV lower than the High Level to
be tested)
High Level Acceptable Range TR Entry
-1.00 V - 1.04 V to -0.96 V 5-6
0 V - 20 mV to +20 mV 5-7
1.00 V 0.96 V to 1.04 V 5-8
2.00 V 1.94 V to 2.06 V 5-9
3.00 V 2.92 V to 3.08 V 5-10
For the 81134A repeat the test for the second channel.