Agilent 81133A / 81134A Performance Verification
Agilent 81133A / 81134A Performance Verification, December 2003
23
Test 5b: Amplitude
Amplitude: 50 mV
pp
to 2 V
pp
Resolution: 10 mV
Accuracy: 2 % of setting ± 20 mV
Digitzing Voltmeter
50 Feedthrough Termination, 0.1 %, 10 W
BNC to Dual Banana plug
SMA Cables 50 Ω, SMA/BNC Adaptor
1 Connect the 50 Feedthrough Termination to the DVM input, using
the BNC to Dual Banana plug
2 Connect the 81133A/81134A channel output to the 50 Feedthrough
Termination, using the SMA/BNC adaptor.
3 Set up the 81133A/81134A with a "zero pattern" as follows.
Equipment needed
Procedure