Agilent Technologies B1500A User's Guide Page 174

  • Download
  • Add to my manuals
  • Print
  • Page
    / 208
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 173
4-18 Agilent B1500 VXIplug&play Driver User’s Guide, Edition 3
Programming Examples for C++ Users
Staircase Sweep Measurement
ret = agb1500_resetTimestamp(vi); /* 39 */
ret = agb1500_force(vi, bulk, agb1500_VF_MODE, 0, 0, 0.1, 0);
ret = agb1500_force(vi, source, agb1500_VF_MODE, 0, 0, 0.1, 0);
ret = agb1500_setIv(vi, gate, agb1500_SWP_VF_SGLLIN, 0, 0, vg, nop, hold, delay,
s_delay, igcomp, pgcomp);
check_err (vi, ret); /* 44 */
ret = agb1500_setSweepSync(vi, drain, agb1500_VF_MODE, 0, 0, vd, idcomp, pdcomp);
check_err (vi, ret);
ret = agb1500_sweepIv(vi, drain, agb1500_IM_MODE, 0, &rep, &sc[0], &md[0], &st[0],
&tm[0]);
check_err (vi, ret); /* 50 */
ret = agb1500_zeroOutput(vi, agb1500_CH_ALL); /* 52 */
ret = agb1500_setSwitch(vi, agb1500_CH_ALL, 0);
check_err (vi, ret); /* 54 */
if ( rep != nop ) { /* 56 */
printf ("%d measurement steps were returned.\nIt must be %d steps.\n", rep, nop);
exit (ret);
} /* 59 */
printf(" Vg (V), Id (mA)\n"); /* 61 */
for (i = 0; i < nop; i++){
printf(" %4.2f, %9.6f \n", sc[i], md[i] * 1000);
} /* 64 */
Line Description
39 Resets time stamp.
40 to 41 Applies voltage to device.
43 Sets the primary sweep source.
46 Sets the synchronous sweep source by using the agb1500_setSweepSync function.
50 Performs staircase sweep measurement by using the agb1500_sweepIv function.
52 Sets the specified port to the zero output state.
53 Disables all ports.
44, 47, 50, and 54 Calls the check_err subprogram (shown in Table 4-1) to check if an error status is
returned for the previous line.
56 to 59 Stops the program execution if the number of returned data is not equal to nop.
61 to 64 Displays the measurement result data.
Page view 173
1 2 ... 169 170 171 172 173 174 175 176 177 178 179 ... 207 208

Comments to this Manuals

No comments