2-64 Agilent B1500 VXIplug&play Driver Users Guide, Edition 3
Driver Functions
agb1500_spotMeas
md[1]: Secondary parameter measurement data (ex: G).
st[n]: Status for the md[n] data.
mon[0]: MFCMU monitor data (AC level monitor data).
mon[1]: MFCMU monitor data (DC bias monitor data).
stmon[n]: Status for the mon[n] data.
where n=0 or 1.
agb1500_spotMeas
This function executes a high speed spot measurement by the specified channel, and
returns the measurement result data, measurement status, and time stamp data.
Syntax ViStatus _VI_FUNC agb1500_spotMeas(ViSession vi, ViInt32 channel,
ViInt32 mode, ViReal64 range, ViPReal64 value, ViPInt32 status, ViPReal64 time);
Parameters vi Instrument handle returned from agb1500_init( ).
channel Slot number of the slot that installs the SMU to be used. See
Table 2-2.
mode Measurement mode. 1 (current) or 2 (voltage).
range Measurement ranging mode. 0 (auto), positive value (limited
auto), or negative value (fixed range). For the available values,
see Table 2-3.
value Measurement data. Returned data.
status Measurement status. Returned data. See “Status Code” on page
2-14. To disable the status data output, set 0 (NULL pointer)
instead.
time Time stamp data (measurement start time). Returned data. To
disable the time stamp data output, set 0 (NULL pointer)
instead.
Page view 79