Agilent Technologies N1022A User's Guide Page 53

  • Download
  • Add to my manuals
  • Print
  • Page
    / 82
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 52
Characteristics and Specifications 3
N2820/1A User’s Guide 53
Table 8 Environmental Characteristics
Item Characteristic
Temperature (operating) Operating: 0° C to 40° C
Non-Operating: –40° C to 70° C
Humidity (operating) Operating: tested at 95% RH @ +40° C
Non-Operating: tested at 90% RH, +65° C
ESD 8 kV HBM
Indoor Use This probe is rated for indoor use only
Table 9 Compatible Oscilloscopes
InfiniiVision Oscilloscopes
3000 X-Series (with software version 2.30 or higher)
4000 X-Series (with software version 3.10 or higher)
Infiniium
a
Oscilloscopes
a Infiniium 90000A and 90000 X/Q-Series oscilloscopes are not compatible
with N2820/1A current probes.
9000A and 9000 H-Series (with software version 4.20 or higher)
Page view 52