74 N2820/1A User’s Guide
5 Performance Verification
Average Gain of Zoomed-In Channel
3 Record the value of +V
in
measured on the multimeter:
+V
in
:
4 Move the multimeter’s leads to the banana plug on the
output of the N1022A/B as shown in Figure 44.
5 Record the value of +V
out
measured on the multimeter:
+V
out
:
Figure 44 Test Setup for Measuring V
out
6 Calculate the positive gain of the zoomed- in (primary)
channel and record the result in the following line: