23
N5250A Configuration Details
N5250A MM PNA system
1
, 10 MHz to 110 GHz, includes:
E8361A MW PNA with the following options:
Configurable test set – Option 014
Extended power range and bias-tees – Option UNL
Frequency-offset mode – Option 080
Reference channel switch – Option 081
IF access – Option H11
N5260A millimeter-wave test set controller with test heads
67 to 110 GHz test heads
1.0 mm combiner assembly
Interconnecting cables
Installation and productivity assistance
Additional options available:
• Millimeter-wave modules with bias-tees -
N5250A Option 017
• Millimeter-wave modules with bias-tees and
port 2 attenuator - N5250A Option 018
Receiver attenuator – Option 016
Time-domain capability – Option 010
Pulsed-RF measurement capability – Option H08
2
Scalar-calibrated converter measurements – Option 082
2
Frequency converter application – Option 083
2
Factory integration of the N5250A system integrates the E8361A
with Option H11 and the N5260A millimeter-wave controller with
test heads. On-site installation is included, and the entire system
carries a full one-year, on-site warranty (where available).
Option Descriptions
Millimeter-wave modules with bias-tees (N5250A Option 017)
Adds 67 GHz bias-tees to the combiner assembly between the
input to the combiner and the 67 GHz coupler. The bias-tees have
tri-axial connectors for force, sense, and ground. Positioning the
bias-tees close to the DUT greatly improves stability for on-wafer
and in-fixture devices. The bias-tees added for this option have a
voltage rating of 40 volts and a maximum of 0.5 amps.
Millimeter-wave modules with bias-tees and port 2
attenuator (N5250A Option 018)
Adds 67 GHz bias-tees to the combiner assembly between the
input to the combiner and the 67 GHz coupler. The bias-tees have
tri-axial connectors for force, sense, and ground. Positioning the
bias-tees close to the DUT greatly improves stability for on-wafer
and in-fixture devices. The bias-tees added for this option
have a voltage rating of 40 volts and a maximum of 0.5 amps
Additionally, Option 018 adds a 25 dB micrometer attenuator to
the port 2 test head.
Banded waveguide solution
In order to assemble a banded waveguide solution, the following
components are needed:
Microwave PNA network analyzer (E8361A or E8362/3/4B)
with the following options:
• IF access - Option H11
• Configurable test set - Option 014
• Extended power range and bias-tees - Option UNL
• Frequency-offset mode - Option 080
• Reference channel switch - Option 081
• Millimeter-wave test set controller (N5260A)
with no options
A set of waveguide modules:
N5260AW15, 50 to 75 GHz
N5260AW12, 60 to 90 GHz
N5260AW10, 75 to 110 GHz
N5260AW08, 90 to 140 GHz
N5260AW06, 110 to 170 GHz
N5260AW05, 140 to 220 GHz
N5260AW03, 220 to 325 GHz
NOTE: To significantly improve system dynamic range above 220 GHz,
Agilent strongly recommends adding two external synthesizers
such as Agilent’s PSG Series signal generators; one for the RF
signal, and one for the LO signal.
For more detailed information, see PNA Millimeter-wave
Technical Overview, literature number 5988-9620EN.
On-wafer applications
For on-wafer applications, Cascade Microtech
3
provides complete
probing systems using the N5250A. These include both new
probing systems and upgrades to existing Cascade Microtech
products. Cascade can also provide on-wafer verification and
probing system training. Once the N5250A system is verified
in coax, Cascade Microtech will verify the system through its
wafer probes.
1. For more detailed information regarding the 110 GHz network analyzer system,
refer to the Agilent Web site: www.agilent.com/find/pna and download the
N5250A Technical Overview, literature number 5988-9620EN.
2. Up to 67 GHz.
3. Cascade Microtech is an Agilent Channel Partner.