1 Characteristics and Specifications
4 U2761A Service Guide
Product Specifications and Characteristics
WAVEFORMS
Standard Sine, Square, Ramp, Triangle, Pulse, DC
Built-in arbitrary Exponential Rise, Exponential Fall, Negative Ramp
WAVEFORM CHARACTERISTICS
Sine
Frequency range 1 μHz to 20 MHz (1 μHz resolution)
Amplitude flatness
1
(relative to 1 kHz)
100 kHz 0.2 dB
100 kHz to 1 MHz 0.35 dB
1 MHz to 20 MHz 0.7 dB
Harmonic distortion
2
Frequency range < 1 Vpp 1 Vpp
DC to 20 kHz –70 dBc –60 dBc
20 kHz to 100 kHz –65 dBc –60 dBc
100 kHz to 1 MHz –50 dBc –45 dBc
1 MHz to 20 MHz –40 dBc –35 dBc
Total harmonic distortion
2
DC to 20 kHz 0.10%
Spurious (Nonharmonic) output
3
DC to 1 MHz –65 dBc
1 MHz to 20 MHz –65 dBc + 6 dB/octave
Phase noise (10 kHz offset) –115 dBc/Hz (Typical)
Square
Frequency range 1 μHz to 20 MHz (1 μHz resolution)
Rise/fall time < 18 ns, 10 to 90% terminated load (50 Ω)
Overshoot < 2%
Variable duty cycle 20% to 80% (up to 10 MHz)
40% to 60% (up to 20 MHz)
Asymmetry (@ 50% duty) 1% of period + 5 ns
Jitter (RMS) > 50 kHz = 1 ns + 100 ppm of period
50 kHz = 10 ns + 100 ppm of period