Agilent Technologies AGILENT U2761A Service Manual Page 35

  • Download
  • Add to my manuals
  • Print
  • Page
    / 67
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 34
Calibration 3
U2761A Service Guide 21
2 Set the U2761A based on the following.
Load impedance: High-Z
Function: DC
3 Set the U2761A to each output listed in the table below.
1 Based upon ±2% of offset setting ±10 mV (High-Z).
4 Compare the measured voltages with the test limits shown
in Table 3- 4.
AC Amplitude (High Impedance) Verification
This procedure verifies the accuracy of the AC output
amplitude at frequencies of 1 kHz and 100 kHz, and
establishes reference measurement for the higher frequency
flatness verification procedures.
1 Set the DMM to measure AC voltage. Connect the DMM to
the U2761A as shown in Figure 3- 3.
2 Set the U2761A based on the following.
Load impedance: High-Z
Offset: 0 V
Table 3 - 4 Configuration for DC offset verification
U2761A Measurement
DC offset Nominal Error
1
–5.000 VDC –5.000 VDC ±0.11 VDC
–2.500 VDC –2.500 VDC ±0.06 VDC
0.000 VDC 0.000 VDC ±0.01 VDC
+2.500 VDC +2.500 VDC ±0.06 VDC
+5.000 VDC +5.000 VDC ±0.11 VDC
Page view 34
1 2 ... 30 31 32 33 34 35 36 37 38 39 40 ... 66 67

Comments to this Manuals

No comments