3 Calibration
30 U2761A Service Guide
Calibration Procedure
Calibration is necessary when the one-year calibration
interval has expired, or whenever the U2761A fails any
verification test. To enter the calibration mode, send the
following SCPI command to the U2761A.
CALibration:SECure:STATe OFF,U2761A
Once the calibration state is unsecured, the U2761A is now
in the calibration mode and setup is 0 by default. All the
calibration setup is predetermined and transparent to the
user. The user is free to perform a full calibration setup
from 1 to 115, or may choose to perform only a particular
group. To perform a particular calibration setup, send the
following SCPI command to U2761A.
CALibration:SETup<1|2|3|...|115>
The current calibration setup can be queried using the
following SCPI command when necessary.
CALibration:SETup?
The U2761A will output a predefined waveform at the output
terminal according to the setup number entered. Perform the
measurements as described in “Internal Timebase
Adjustment” on page 31 until “Amplitude Flatness (+20 dB
Range) Adjustment” on page 41, and enter the measured
values to the U2761A using the SCPI commands as follows.
CALibration:VALue <value>
The calibration value must be entered at the magnitude of
the units shown in the configuration tables accordingly. If
the user accidentally enters a wrong value, simply reenter a
correct value under the same calibration setup. Proceed to
the next calibration setup by entering a new setup number
once the calibration value is entered.