Calibration 3
U2761A Service Guide 19
2 Set the U2761A to the output shown in the table below
and measure the output frequency. Make sure that the
U2761A output is enabled.
1 Based upon ±8 ppm for one year.
3 Compare the measured frequency with the test limit
shown in Table 3- 2.
Output Impedance Verification
The U2761A stores the calibration constant for the output
impedance. The output impedance measurement is necessary
to calculate the correct output amplitude when the load
impedance is not at High- Z termination. The output
impedance is measured at three different output paths.
1 Set the DMM to measure offset-compensated, four-wire
ohms. Set the DMM to use 100 NPLC integration. Connect
the DMM to the U2761A as shown in the figure below.
Figure 3-2 U2761A to DMM four-wire connection
Table 3 - 2 Configuration for timebase verification
U2761A Measurement
Function Amplitude Frequency Nominal Error
1
Sine 1.00 Vpp 10.0000 MHz 10.0000 MHz ±80 Hz
Page view 32