MEASurement Instructions
READ[1|2] Commands
Agilent E4418B/E4419B Programming Guide 2-35
READ[1|2] Commands
The READ? commands are most commonly used with the CONFigure
command to cause a new power measurement to be taken and the result
returned to the output buffer. The format of the return is <NR3>, for
example, -6.98970004E-001.
• For the Agilent E4418B the READ? query is equivalent to:
ABORt
INITiate
FETCh?
• For the Agilent E4419B carrying out a single channel
measurement the READ? queries are equivalent to:
ABORt1
INITiate1
FETCh1?
or
ABORt2
INITiate2
FETCh2?
• For the Agilent E4419B carrying out a difference measurement
the READ:DIFFerence? queries are equivalent to:
ABORt1and
ABORt2
INITiate1
INITiate2
FETCh:DIFFerence?
• For the Agilent E4419B carrying out a ratio measurement the
READ:RATio? queries are equivalent to:
ABORt1
ABORt2
INITiate1
INITiate2
FETCh:RATio?
prog.book Page 35 Thursday, June 7, 2001 2:55 PM