MEMory Subsystem
MEMory:CATalog[:ALL]?
7-4 Agilent E4418B/E4419B Programming Guide
MEMory:CATalog[:ALL]?
This command is used to list the stored sensor calibration tables,
frequency dependent offset tables and save/recall registers.
The power meter returns the data in the form of two numeric parameters
and as many strings as there are stored tables and save/recall registers.
<numeric_value>,<numeric_value>{,<string>}
The first numeric parameter indicates the amount of memory, in
bytes, used for the storage of tables and registers.
The second numeric parameter indicates the memory, in bytes,
available for the storage of tables and registers.
Each string parameter returned indicates the name, type and size
of a stored table or save/recall register:
<string>,<type>,<size>
<string> indicates the name of the table or save/recall
register.
<type> indicates TABL for sensor calibration and frequency
dependent offset tables, or STAT for a save/recall register.
<size> indicates the size of the table or save/recall register
in bytes.
For example, a sample of a response may look like:
1178,26230,"DEFAULT,TABL,14","HP8481A,TABL,116",
"HP8482A,TABL,74",..........."State0,STAT,1619",
"State1,STAT,1619","State2,STAT,1619" ...........
prog.book Page 4 Thursday, June 7, 2001 2:55 PM