CALibrate:SAVE
This command saves any new calibration constants after a calibration procedure has been completed in
nonvolatile memory. If CALibrate:STATe OFF is programmed without a CALibrate:SAVE, the previous
calibration constants are restored..
Command Syntax
CALibrate:SAVE
Parameters
None
Examples
CAL:SAVE
Related Commands
CAL:PASS CAL:STAT
CALibrate:STATe
This command enables and disables calibration mode. The calibration mode must be enabled before the
dc source will accept any other calibration commands.
The first parameter specifies the enabled or disabled state. The second parameter is the password. A
password is required if calibration mode is being enabled and the existing password is not 0. If the
password is not entered or is incorrect, an error is generated and the calibration mode remains disabled.
The query returns only the state, not the password.
NOTE: Whenever the calibration state is changed from enabled to disabled, any new calibration
constants are lost unless they have been stored with CALibrate:SAVE.
Command Syntax
CALibrate:STATe <bool>[,<NRf>]
Parameters
0 | OFF | 1 | ON [,<password>]
*RST Value
OFF
Examples
CAL:STAT 1,3280 CAL:STAT OFF
Query Syntax
CALibrate:STATe?
Returned Parameters
<NR1>
Related Commands
CAL:PASS CAL:SAVE *RST
CALibrate:VOLTage
This command initiates the calibration of the output voltage and the voltage measurement circuit.
Command Syntax
CALibrate:VOLTage (@<channel>)
Parameters
None
Examples
CAL:VOLT (@1)