MEASure:ARRay:CURRent?
MEASure:ARRay:VOLTage?
These queries initiate and trigger a measurement and return an array containing either the digitized output
current in amperes or output voltage in volts. The output voltage or current is digitized whenever a
measurement command is sent or an acquisition trigger occurs. The time interval is set by
SENSe:SWEep:TINTerval. The position of the trigger relative to the beginning of the data buffer is
determined by SENSe:SWEep:OFFSet. The number of points returned is set by SENSe:SWEep:POINts.
NOTE: You can specify an optional maximum value parameter. The dc source will select the
proper current range to measure the maximum current.
Query Syntax
MEASure:ARRay:CURRent[:DC]? [<maximum value>,] (@<channel list>)
MEASure:ARRay:VOLTage[:DC]? (@<channel list>)
Parameters
None
Examples
MEAS:ARR:CURR? 0.1,(@1) MEAS:ARR:VOLT? (@1)
Returned Parameters
<NR3>[,<NR3>]
Related Commands
SENS:SWE:TINT SENS:SWE:OFFS SENS:SWE:POIN
MEASure:CURRent?
MEASure:VOLTage?
These queries initiate and trigger a measurement and return either the output current in amperes or
output voltage in volts. The total measurement time is specified by SENSe:SWEep:NPLCycles.
NOTE: You can specify an optional maximum value parameter. This lets you use a different
current range for a single measurement without having to change current ranges.
Query Syntax
MEASure[:SCALar]:CURRent[:DC]? [<maximum value>,] (@<channel list>)
MEASure[:SCALar]:VOLTage[:DC]? (@<channel list>)
Parameters
None
Examples
MEAS:CURR? 0.1,(@1) MEAS:VOLT? (@1)
Returned Parameters
<NR3> [,<NR3>]
Related Commands
FETC:VOLT? FETC:CURR? SENS:SWE:NPLC
SENSe:CURRent:RANGe
This command selects one of the following dc current measurement ranges based on the value entered:
0.5 A
Enter values greater than 0.015A
15 mA
Enter values greater than 0.0005A up to 0.015A
0.5 mA
Enter values less than and up to 0.0005A
The programmed value must be the maximum current that you expect to measure. Crossover values are
0.5 mA and 15 mA respectively. When queried, the returned value is the maximum dc current that can be
measured on the range that is presently set.
Command Syntax
SENSe:CURRent[:DC]:RANGe[:UPPer] <Nrf>, (@<channel list>)
Parameters
The maximum current that you expect to measure (see table 6-3)
Unit
A (amperes)
*RST Value
0.5 A
Examples
SENS:CURR:RANG 0.4,(@1)
Query Syntax
SENSe:CURRent:RANGe? (@<channel list>)
Returned Parameters
<NR3>