10 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20/40 GHz
1 Security Features
Removing Sensitive Data from Memory
This section describes several security functions that can be used to remove sensitive
data stored in the N8211A performance analog upconverter and N8212A performance
vector upconverter when moving them from a secure development environment.
NOTE
All security functions described in this section have an equivalent SCPI command for
remote operation. For more information about the SCPI commands available for removing
sensitive data from memory, refer to “SCPI Commands" on page 15.
Declassification
To declassify the instrument in compliance with Department of Defense (DoD)
standards, perform the following steps:
1 Send the SCPI command SYSTem:SECurity:LEVel SANitize
2 Send the SCPI command SYSTem:SECurity:LEVel:STATe ON
3 Send the SCPI command SYSTem:SECurity:LEVel:STATe OFF
4 Remove the battery, leave it out of the instrument for one minute, and replace it.
Refer to “Battery Removal/Replacement" on page 23 for details.