14 Agilent N8211A/N8212A Performance Upconverter Synthetic Instrument Module, 250 kHz to 20/40 GHz
1 Security Features
User IQ Cal File (Vector Models Only)
When a user-defined IQ calibration has been performed, the cal file data is removed by
setting the cal file to default, as follows:
Send these commands to the N8211A or N8212A upconverter:
CAL:IQ:DEF
CAL:WBIQ:DEF
If Your Instrument is Not Functioning
If the instrument is not functioning and you are unable to use the security functions, you
may physically remove the processor board and hard disk, if installed, from the
instrument. Once these assemblies are removed, proceed as follows:
For removal and replacement procedures, refer to the procedure at the end of this
document.
Processor Board
Either
Discard the processor board and send the instrument to a repair facility. A new
processor board will be installed and the instrument will be repaired and calibrated. If
the instrument is still under warranty, you will not be charged for the new processor
board.
or
If you have another working instrument, install the processor board into that
instrument and erase the memory. Then reinstall the processor board back into the
non-working instrument and send it to a repair facility for repair and calibration. If
you discover that the processor board does not function in the working instrument,
discard the processor board and note that it caused the instrument failure on the repair
order. If the instrument is still under warranty, you will not be charged for the new
processor board.