Troubleshooting 5
16800 Series Portable Logic Analyzers Service Guide 103
e Drag the scroll bar all the way to the left and ensure
that the activity indicator shows activity on clock 1.
3 Set the sampling mode.
a Select the Sampling tab of the Analyzer Setup dialog.
b Select State - Synchronous Sampling.
c Set the Trigger Position to 100% Poststore.
d Set the Acquisition Depth to 128K.
e Clear the Timing Zoom check box to turn Timing Zoom
off.
f Ensure that the sampling speed is set to 250 MHz in
the Sampling Options box.