Agilent Technologies 8133A Service Manual Page 108

  • Download
  • Add to my manuals
  • Print
  • Page
    / 188
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 107
108 16800 Series Portable Logic Analyzers Service Guide
5 Troubleshooting
10 Click Run, and ensure that an eye is found for each bit.
11 Click OK to close the Thresholds and Sample Positions
dialog.
12 Click OK to close the Analyzer Setup dialog.
13 Switch to the Listing window and run the logic analyzer.
14 Examine the listing. If the listing shows that the data
alternates between 0 AAAA and 1 5555, then the probe
and cable pass the test.
If the listing does not look similar to the figure, then
there is a possible problem with the cable or probe tip
assembly. Causes for cable test failures include:
Open channel.
Channel shorted to a neighboring channel.
Channel shorted to either ground or a supply voltage.
Page view 107
1 2 ... 103 104 105 106 107 108 109 110 111 112 113 ... 187 188

Comments to this Manuals

No comments