Replacing Assemblies 6
16800 Series Portable Logic Analyzers Service Guide 131
4 If you have the contrast filter, it sets in a groove in the front
panel.
5 If you have the touch screen option, gently pry the touch
screen from the LCD. It is held on with double-sided tape.
6 Reverse this procedure to replace the display assembly.
CAUTION
Handle the touch screen and LCD display by the edges only. If
necessary, clean with alcohol and a lint-free wipe or remove any
airborne contamination with ionized air.