136 16800 Series Portable Logic Analyzers Service Guide
6 Replacing Assemblies
4 Connect the measurement cable(s).
a Align the logic analyzer cable end connector with the
circuit board connector and gently apply pressure to
seat the cable.
b Install the Torx T10 screws (two per cable) and tighten
to 5 in/lb to secure the cable to the back panel of the
card.
Back panel(s)
T-10 screws
Pods 7 & 8
Pods 1 & 2
Pods 11 & 12
Pods 5 & 6
Pods 9 & 10
Pods 3 & 4