U2701A/U2702A
Performance specifications
[1]
Vertical system: oscilloscope channels
Bandwidth (–3 dB) U2701A: DC to 100 MHz
U2702A: DC to 200 MHz
Scope channel triggering
Trigger sensitivity < 10 mV/div: greater of 1 div or 5mV; ≥10 mV/div: 0.6 div
Performance characteristics
[2]
Acquisition: oscilloscope channels
Real time sample rate
2 channels interleaved
Each channel
1 GSa/s
500 MSa/s
Standard memory depth
2 channels interleaved
Each channel
Normal Single-shot
32 Mpts 64 Mpts
16 Mpts 32 Mpts
Vertical resolution 8 bits
Peak detection Yes
Averaging Any number from 1 to 999
Filter Sin(x)/x interpolation for time base 1 ns to 100 ns
Sweep modes Auto, normal, single
Vertical system: oscilloscope channels
Scope channels U2701A/U2702A: Ch 1 and Ch 2 simultaneous acquisition
AC coupled U2701A: 3.5 Hz to 100 MHz
U2702A: 3.5 Hz to 200 MHz
Calculated rise time (= 0.35/bandwidth) U2701A: 3.5 ns
U2702A: 1.75 ns
Single-shot bandwidth U2701A: 100 MHz
U2702A: 200 MHz
Range 2 mV/div to 5 V/div (1 MΩ)
Maximum input
[3]
CAT I 30 Vrms, 42 Vpk
Offset range
±4 div
Example: ±8 mV on 2 mV/div; ±20 V on 5 V/div
Dynamic range ±4 div
Input impedance 1 MΩ: ≈ 16 pF
Coupling AC, DC, Ground
BW limit ≈ 25 MHz
Standard probes 10:1 Passive probe 150 MHz 1.2 m
10:1 Passive probe 300 MHz 1.2 m
ESD tolerance ±2 kV
Noise peak-to-peak 3 mVpp
DC vertical offset accuracy ≤ 200 mV/div: ±0.1 div ±2.0 mV ±0.5% offset value;
> 200 mV/div: ±0.1 div ±2.0 mV ±1.5% offset value
DC vertical gain accuracy ±4.0% of full scale
[1]
All specifications are warranted. specifications are valid after a 30-minute warm-up and within ±100 ºC of last calibration temperature.
[2]
All characteristics are typical performance values and are not warranted. Characteristics are valid after a 30-minute warm-up period and within
±10 ºC of last calibration temperature.
[3]
Under standalone use, you are only allowed to measure up to CAT I 30 Vrms. For high-voltage measurement up to CAT I 300 Vrms, you must install the
L-Mount kit on the U2701A/U2702A before plugging it into the product chassis. Ensure that the L-Mount kit installed on your modular oscilloscope is
screwed to the product chassis to ensure proper chassis grounding. Note that you are required to use the provided 10:1 probes (N2862A/N2863A)
for high-voltage measurements to avoid damaging your instrument.