6 Specifications and Characteristics
84 U3401A User’s and Service Guide
DC Specifications
Table 6 - 1 DCV resolution, full scale reading and accuracy [±(% of reading + count)]
Function Range Resolution
Maximum
reading
Accuracy
(One year; 23°C ±
5 °C)
Test Current
Typical input
impedance
[2]
Burden voltage
[3]
Voltage
500 mV 0.01 mV 510.00 0.02% + 4 10.0 MΩ
5 V 0.0001 V 5.1000 0.02% + 4 11.1 MΩ
50 V 0.001 V 51.000 0.02% + 4 10.1 MΩ
500 V 0.01 V 510.00 0.02% + 4 10.0 MΩ
1000 V 0.1 V
1200.0
[1]
0.02% + 4 10.0 MΩ
Current
500 μA
0.01 μA 510.00 0.05% + 5 < 0.06 V
5 mA
0.0001 mA 5.1000 0.05% + 4 < 0.6 V
50 mA
0.001 mA 51.000 0.05% + 4 < 0.08 V
500 mA
0.01 mA 510.00 0.05% + 4 < 0.8 V
5 A
0.0001 A 5.1000 0.25% + 5 < 0.3 V
10 A
0.001 A
20.000
[4]
0.25% + 5 < 0.6 V
Resistance
/ Continuity
[5]
500 Ω
0.01 Ω 510.00
0.1% + 5
[6]
0.5 mA
5 kΩ
0.0001 kΩ 5.1000
0.1% + 3
[6]
0.45 mA
50 kΩ
0.001 kΩ 51.000 0.1% + 3 45 μA—
500 kΩ
0.01 kΩ 510.00 0.1% + 3 4.5 μA—
5 MΩ
0.0001 MΩ 5.1000 0.1% + 3 450 nA
50 MΩ
0.001 MΩ 51.000 0.3% + 3 45 nA
Diode Test /
Continuity
2.3 V
0.0001 V 2.3000 V 0.05% + 5
Notes:
1 In 1000 V range, 1200 V is readable with audio warning.
2 Input impedance is in paralleled with capacitance <100 pF.
3 Typical at full scale reading and voltage across the input terminals.
4 In 10 A range, >10-20 Adc is readable for 20 seconds maximum with audio warning.
5 In order to eliminate the noise interference, which might be induced to the test leads, it is recommended to use a shielded
test cable for measuring resistance above 500 KΩ.
6 Use relative operation. If relative operation is not used, add 0.2 Ω additional error.