Agilent 5500 SPM User’s Guide 191
container; therefore, the sample must be large enough for the O-ring to
seat.
Figure 125 Li
quid cell components
Figure 126 Liquid cell mounted on standard sample plate
Liquid Cell with Standard Sample Plate
One challenge with using the liquid cell is to locate the region of interest
through the liquid. It is typically easier to first locate the dry sample and
then to add the liquid, as follows:
1 Prepare and place the sample on the sample plate.
2 Place the sample plate on the microscope.
3 Using the Close switch on the Head Electronics Box, and watching
the video system, roughly approach the sample such that the tip and