Agilent 5500 SPM User’s Guide 67
Two-Piece Nose Assembly
The two-piece nose assembly was designed to simplify the process of
inserting a probe through the use of an assembly fixture (Figure 41).
The nose assembly consists of a body, which inserts into the scanner,
and a flat, stainless steel disk which holds the cantilever. The disk is
held to the body magnetically and can be separated by holding the disk
at its edges and gently pulling it from the body.
Figure 41 Two-piece nose assembly, removal tool and assembly fixture.
The nose assembly disk is shown on the fixture; the body is shown to
its left.
Inserting the Body of the Two-Piece Nose Assembly
As with the one-piece nose assembly, the body of the two-piece
assembly is held in the scanner by an O-ring. To insert the body, first
place the scanner in the scanner mounting fixture (Figure 32 on
CAUTION
The retainer spring can snap back with enough force to damage the
probe, so be sure to release the spring slowly and gently.