Agilent Technologies B1500A User's Guide Page 88

  • Download
  • Add to my manuals
  • Print
  • Page
    / 202
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 87
3- 10 Agilent B1500A User’s Guide, Edition 13
Installation
Inspection and Installation
To Change GPIB Address
When this instrument is shipped from the factory, the GPIB address is set to 17. To
change the GPIB address, perform the following procedure.
1. If the EasyEXPERT software is running, terminate it as shown below:
a. Select the menu function File > Exit on the EasyEXPERT main screen.
b. Click [x] at the upper right corner of the Start EasyEXPERT button.
2. Select All Programs > Agilent IO Libraries Suite > Agilent Connection Expert
from the Start menu. Agilent Connection Expert window appears.
3. At the Instrument I/O on this PC area, select GPIB0, and click the Change
Properties... button. Agilent 82350 PCI GPIB Interface - GPIB0 window
appears.
4. Change the GPIB Address value.
5. Remove the check from the Auto-discover instruments connected to this
interface box.
6. Click the OK button on Agilent 82350 PCI GPIB Interface - GPIB0 window.
7. If the Reboot Required dialog box is displayed, click the Reboot Now button,
and reboot the instrument.
Page view 87
1 2 ... 83 84 85 86 87 88 89 90 91 92 93 ... 201 202

Comments to this Manuals

No comments