Agilent Technologies B1500A User's Guide Page 98

  • Download
  • Add to my manuals
  • Print
  • Page
    / 202
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 97
3- 20 Agilent B1500A User’s Guide, Edition 13
Installation
Installing Accessories
Figure 3-2 To Connect ASU
Figure 3-3 To Set Shorting Bar
To connect other
instrument to AUX
To connect the instrument other than the MFCMU or the 4284A/E4980A:
1. Turn the instrument off.
2. Connect the BNC cable from the ASU CMU-cur/AUX In terminal to the
instrument. Prepare an adapter if the instrument’s input/output connector is not
BNC.
&085HWXUQ$8;&RPPRQ
7R'87
7R&08
)RUFH
+S
+F
'VXE
7R608
7R&08
)RUFH
/S
/F
'VXE
7R608
7R'87
&085HWXUQ$8;&RPPRQ
)RUFH&08$8;
6HQVH
&08FXU$8;,Q
&08SRW
)RUFH
)URP608
$68
&08FXU$8;,Q
&08SRW
)RUFH
)URP608
$68
)RUFH&08$8;
6HQVH
CMU-cur/AUX In
SMU Force
Shorting bar
AUX
SMU Force
b / c
5 / 8
6 / 9
Page view 97
1 2 ... 93 94 95 96 97 98 99 100 101 102 103 ... 201 202