Function Switching
The purpose of the Function Switching section (schematic shown on
page 9-9) is to connect the
Input HI terminal to the various measuring
functions. This is accomplished through
K101, K102, K103, and K104.
In addition, the connections of the 4-wire ohms
HI Sense and LO Sense
inputs are shown. Shunt selection (ranging) and voltage sensing are also
shown for the current function. The table below shows the state of each
relay for each measuring function. Relay
K101 opens momentarily during
each range or function change. All relay coils are driven from
U150.
Function K101 K102 K103 K104 Sense at:
0.1 V 10 Vdc
100 V
1000 Vdc
2-Wire Ohms
4-Wire Ohms
AC Voltage
Frequency/Period
3 A, 1A DC I
100 mA, 10 mA DC I
3A, 1A AC I
Closed
Closed
Closed
Closed
Closed
Closed
Closed
Closed
Closed
Set
Set
Reset
Reset
Set
Set
Reset
Set
Reset
Set
Reset
Set
Set
Reset
Reset
Set
Set
Set
Set [1]
Set
Reset [2]
Reset [2]
Reset
Reset
Reset
Reset
Set
U101-5
U102-12
U101-5
[3]
AC_IN
AC_IN
U101-10
U101-10
AC_IN
[1] K104 will be reset when input resistance is selected to >10,000 M
through the menu.
[2] K104 will be set for the 100 M
range.
[3] Configurations shown are for the current source output (HI) terminal. The measurement
sense is accomplished through the Sense HI / Sense LO terminals.
5
Chapter 5 Theory of Operation
Function Switching
95