Agilent Calibration Services
When your multimeter is due for calibration, contact your local Agilent
Service Center for a low-cost recalibration. The 34401A Multimeter is
supported on automated calibration systems which allow Agilent to
provide this service at competitive prices. Calibrations to MIL-STD-45662
are also available at competitive prices.
Calibration Interval
The multimeter should be calibrated on a regular interval determined
by the measurement accuracy requirements of your application.
A 90-day interval is recommended for the most demanding applications,
while a 1-year or 2-year interval may be adequate for less demanding
applications. Agilent does not recommend extending calibration
intervals beyond 2 years for any application.
Whatever calibration interval you select, Agilent recommends that
complete re-adjustment should always be performed at the calibration
interval. This will increase your confidence that the Agilent 34401A will
remain within specification for the next calibration interval. This
criteria for re-adjustment provides the best measure of the multimeter’s
long-term stability. Performance data measured using this method can
easily be used to extend future calibration intervals.
Time Required for Calibration
The Agilent 34401A can be automatically calibrated under computer
control. With computer control you can perform the complete calibration
procedure and performance verification tests in less than 20 minutes.
Manual calibrations using a multi-function calibrator will take
approximately 40 minutes.