Agilent Technologies B1500A Service Manual Page 109

  • Download
  • Add to my manuals
  • Print
  • Page
    / 110
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 108
Module 6
Low Current Measurement
6-37
Negative sweep.
Accumulation mode.
Wait for 5 V initial step
FOWLER-NORDHEIM PLOT
Measurement Page
A 3 second delay at the beginning of the measurement will be required. The SMU must initially step
from 0 to -5 V, a very large step. Fully guarded probes to within 2 mm of probe tip and fully guarded
chuck eliminated the need for delays at every measurement point. The only wait will be the initial 3
second hold time.
Page view 108
1 2 ... 104 105 106 107 108 109 110

Comments to this Manuals

No comments