Page view 64
Module 5
Basic Measurement
5-45
Non-Kelvin Connection
Jumper to eliminate cable
resistance in force line
Insulator
COMMON
FORCE
SENSE
Shielding Box
Connector Plate
16495H-002
or
16495J-002
GUARD
GUARD
COMMON
FORCE,SENSE Line
GUARD Line
Kelvin Triaxial Cable
16493K
Probe Cable (Coaxial)
Prober
DUT
S
M
U
F
S
F
S
Connector Plate
This is a "hybrid" Kelvin connection. Cable resistance from SMU is eliminated by shorting FORCE
and SENSE lines inside the prober. A short line to the probe makes the non-Kelvin connection,
simplifying hookup.