Module 6
Low Current Measurement
6-5
9 Low loss triaxial connection (no matrix card)
9 Guard to tip of probe (no bare needle or probe cards)
9 No air flowing near probe tip (including dry nitrogen)
9 No hot chuck
9 Chuck isolated from ground (10^13 ohms)
9 Chuck guarded for substrate measurements
9 Shielded probe station
9 No vibration; cables stationary
9 Microscope light, motors, etc. off
Clean Probing Environment
What is Required?
This check list covers most sources of noise or stray capacitance. In a "clean" probing environment
the B1500A can be used with short integration time and no delay time between steps.