114 34980A Service Guide
5 Disassembly and Repair
Electrostatic Discharge (ESD) Precautions
Almost all electrical components can be damaged by electrostatic discharge
(ESD) during handling. Component damage can occur at electrostatic
discharge voltages as low as 50 volts.
The following guidelines will help prevent ESD damage when servicing the
instrument or one of the plug- in modules.
Disassembly the instrument only in a static- free work area.
Use a conductive work area to dissipate any static charge.
Use a conductive wrist strap to dissipate static charge accumulation.
Minimize handling.
Keep replacement parts in their original static- free packaging.
Remove all plastic, foam, vinyl, paper, and other static- generating
materials from the immediate work area.
Use only anti- static solder suckers.
Surface Mount Repair
Surface mount components should only be removed using soldering irons
or desoldering stations expressly designed for surface mount components.
Use of conventional solder removal equipment will almost always results
in permanent damage to the printed circuit board.
Tools Required
The following tools are required for basic disassembly.
T20 Torx driver
3/16” nut driver (for rear panel connectors)
9/32” nut driver (for GP- IB connector)
#2 Pozidrive (for fan)