Specifications 2
34980A Service Guide 39
Interpreting Internal DMM Specifications
This section is provided to give you a better understanding of the
terminology used and will help you interpret the internal DMM’s
specifications.
Number of Digits and Overrange The “number of digits” specification is
the most fundamental, and sometimes, the most confusing characteristic of
a instrument. The number of digits is equal to the maximum number of
“9’s” the instrument can measure or display. This indicates the number of
full digits. Most instruments have the ability to overrange and add a
partial or “½” digit.
For example, the internal DMM can measure 9.99999 Vdc on the 10 V
range. This represents six full digits of resolution. The internal DMM can
also overrange on the 10 V range and measure up to a maximum of
12.00000 Vdc. This corresponds to a 6½- digit measurement with 20%
overrange capability.
Sensitivity Sensitivity is the minimum level that the instrument can
detect for a given measurement. Sensitivity defines the ability of the
instrument to respond to small changes in the input level. For example,
suppose you are monitoring a 1 mVdc signal and you want to adjust the
level to within ±1
V. To be able to respond to an adjustment this small,
this measurement would require a instrument with a sensitivity of at least
1
V. You could use a 6½- digit instrument if it has a 1 Vdc or smaller
range. You could also use a 4½- digit instrument with a 10 mVdc range.
For ac voltage and ac current measurements, note that the smallest value
that can be measured is different from the sensitivity. For the internal
DMM, these functions are specified to measure down to 1% of the selected
range. For example, the internal DMM can measure down to 1 mV on the
100 mV range.
Resolution Resolution is the numeric ratio of the maximum displayed
value divided by the minimum displayed value on a selected range.
Resolution is often expressed in percent, parts- per- million (ppm), counts,
or bits. For example, a 6½- digit instrument with 20% overrange capability
can display a measurement with up to 1,200,000 counts of resolution. This
corresponds to about 0.0001% (1 ppm) of full scale, or 21 bits including
the sign bit. All four specifications are equivalent.