72 34980A Service Guide
3 Calibration Procedures
34952A Multifunction Module
The only calibration constants are for the two DAC outputs on the 34952A
Multifunction Module. These outputs can be measured using an external
voltmeter, or with a test fixture such as the one shown below, using the
internal DMM via the ABus connector on the instrument’s rear panel.
34952A Verification
Analog Output Verification Test This procedure is used to check the
calibration of the analog outputs on the 34952A Multifunction Module.
Verification checks are performed only for those output values with unique
calibration constants.
1 Use the internal DMM to measure the output of each channel. You will
need to physically move the input connections on the 34952A module to
channel 6 and then channel 7.
2 For each analog output (channel 6 and channel 7), set each output
value in the table on the next page. Compare measurement results to
the appropriate test limits shown in the table.
NOTE
It is not necessary to test the voltage output at the full rated 10 mA load.