Disassembly and Repair 5
34980A Service Guide 115
Basic Disassembly
Observe the electrostatic discharge precautions given on page 114.
1 Remove the power cable from the unit. If attached, remove the analog
bus connector.
a If desired, you may also remove the feet by lifting the tab on each
foot and sliding the foot toward the rear of the instrument.
2 Remove all plug- in modules.
3 Using a T- 20 Torx driver, loosen the five captive screws in the rear
bezel and remove the bezel. The metal cover will now slide off.