Calibration Procedures 3
34980A Service Guide 71
Analog Output Adjustment
Install the 34951A module in the mainframe and allow a 1 hour warm- up
before performing these procedures.
This adjustment procedure sets a zero adjustment and a gain adjustment
constant for each DAC output. You must perform all the adjustments on
one analog output channel before adjusting the other analog output
channel.
1 Install the module(s) in the instrument. Remove any inputs from the
ABus connector.
2 Set the calibration security for the adjustment mode you desire, see
page 66.
3 Send the CALibration:MODule? <slot> command to begin the
auto- calibration procedure for all channels on the module in
<slot>.
You may use the
CALibration:MODule? ALL command to perform the
auto- calibration on all channels for all modules installed in the
mainframe. The complete procedure may take up to 1 minute per
module. Be sure to set the remote command time- out value
appropriately.
4 The command returns a value of “+0” if the calibration is successful.
A returned value of “+1” indicates a calibration failure.
CAUTION
Remove any ABus connector before performing this procedure.
NOTE
Following the auto-calibration procedure, the DMM is left in its reset
state (DCV).