Lab Exercise: Running a Sequence
Sequence Summary Reports
104
Sequence Summary Reports
The ChemStation can print a variety of standard reports for individual sample
analyses as directed in the Specify Report panel. These styles are stored as part
of the method and are printed at the end of each analysis if desired. Sequence
summary reporting is an additional, optional way of reporting. You may calculate
and report parameters across a number of different analyses, for instance, you
may report the standard deviation of the retention time for calibration standards.
These calculations are useful, for example, to test instrument performance or the
stability of a new method. You will add a sequence summary report to this
sequence. This type of report will print at the end of the entire sequence.
1) To begin setting up your sequence summary report, select Sequence Output
from the Sequence menu or from the sequence tray setup tool.
2) Select the Print Sequence Summary Report option along with the Report to
Printer and Print individual reports for each run as well, options.
Question:
If you had not selected Print individual reports for each run as well, would you
receive an individual report after each sample had been run?
________________________________________________________________
3) Specify the destination for individual reports as the Printer, as specified
here, not as specified with each method.
4) Click on the Setup... button. Select items 4-9. For items 7 and 8, select
Extended Statistic. For item 9, select Sample Summary. OK this dialog
box and the Sequence Output dialog box.
5) To specify the Extended Statistics, go to the Sequence menu and select
Extended Statistics…. Select any item you would like calculated by double
clicking the Active box. Note that you can place limits in the field to the right
that will be marked in the report if exceeded. OK this dialog box.
6) Resave the sequence.